What is Phaco Emulsification?
Tell me about the History of Phaco Emulsification.
How Does a Phaco Work?
What is the Phaco Procedure?
Phaco Handpiece.
What are the Phaco Handpiece Components?
Tell me about the Phaco Piezoelectirc crystal.
What are common Reasons for handpiece failure?
Phaco Handpeice Repair Procedure.
Phaco repair considerations.
Service Options?
- Surgical cataract removal procedure
- Number 1 therapeutic surgery for Americans over 65 years of age.
- 1.4 Million procedures performed each year
Tell me about the History of Phaco Emulsification:
- Definition: Phaco lens shaped, relating to a lens. From the Greek phakos, lentil (lens), anything shaped like a lentil
- Invented nearly 40 years ago by Dr. Charles Kelman, M.D.
- He received a $280,000 grant and a 3 year time limit
- The idea of using ultrasound cam to him as he sat in the dentist’s chair.
- Operates on the principle of the Piezoelectric Effect
- Piezoelectric Effect discovered by Pierre & Jacques Curie in the 1880’s
- When Crystals are subjected to an electric field, they expand and contract in direct proportion to that electrical field.
- Converting electrical energy into mechanical energy.
What is the Phaco Procedure: Top of Page
- Handpeice primed and tuned
- Small incision in the eye via diamond knife
- Handpiece tip inserted under the cornea
- Phaco machine powers crystals in the handpiece
- Handpiece tip vibrates ultrasonically thousands of times per second to break up (emulsify) the cataract
- Cataract aspirated and flushed with fluid solution
- Man-made Intra-Ocular Lens inserted
- Microprocessor controlled instrument
- Has capability of irrigation, aspiration, emulsification, cautery, and cut modes of operation
- Functions can be controlled by foot pedal, remote control, or on the console.
Phaco Handpiece: Top of Page
- A Slim, lightweight, piezo-electric, air-cooled, autoclavable
- Instrument approximately the size of a magic marker
What are the Phaco Handpiece Components: Top of Page
- Distal Tip
- Stainless steel or Titanium casing/shell
- Cord strain relief
- Aspiration port
- Irrigation port
- Serial #, Make and Model etching
Tell me about the Phaco Piezoelectirc crystal: Top of Page
- Pair of ceramic crystals which vary in size resembling thick washers. When electrified the tip vibrates longitudinally (like a jackhammer) thousands of times per second. Crystals are extremely sensitive to changes in temperature. There are no moving parts in a Handpiece.
What are common Reasons for handpiece failure: Top of Page
- 500 Uses in general will lead to some kind of failure
- Crystals cracked
- Cord (wire) damage
- Connector damage
- Tip damage
- Jammed Needles
- Clogged aspiration port
- Sterilization is the primary cause of failure even under normal use
- Repeated exposure to heat and pressure causes the crystals to depolarize, or causes breaks in the mold/seals which allows moisture intrusion.
- Moisture intrusion will cause electrical failure
- Always follow the manufacturers sterilization and cooling instructions.
Phaco Handpeice Repair Procedure: Top of Page
- Disassemble Handpeice and inspect
- Discard cable, connector, crystals, electrodes and o-rings
- Inspect and clean salvaged components ultrasonically:
Shell, header assembly, horn, etc. Top of Page
- Complete new subassemblies and test each. Cable assembly, transducer stack assembly and cable stack assembly and final assembly.
- Final test, autoclave and retest
- Ship to customer.
Phaco repair considerations: Top of Page
- Most of the time a handpiece must be disassembled to determine which components have failed. The machine does not provide a failure analysis on the handpiece. Only external damage (cord and tip damage) and certain connector problems can be detected prior to disassembly.
- Manufacturers do not trouble-shoot or do minor repairs. They send out exchange handpieces.
- Manufacturer does not repair. Handpieces are replaced.
- Standard warranty is 90 days on component failure or on workmanship.
Service Options: Top of Page
- Send in your handpiece and receive a quote the next day.
- Approved repairs ship within 48 hours. (Ozil and Neosonix require and extra 24 hours)
- Emergency exchanges are available. Give us a call. We overnight a handpeice to you and you send us your damaged handpiece the next day.
- No loaners are available.